MAUDE Adverse Event Report: SPACELABS HEALTHCARE XHIBIT CENTRAL STATION; CENTRAL STATION REMOTE MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SPACELABS HEALTHCARE XHIBIT CENTRAL STATION; CENTRAL STATION REMOTE MONITOR Back to Search Results
Model Number 96102
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/04/2025
Type of Reportable Event Malfunction
Event or Problem Description
The customer contacted spacelabs healthcare to report that two displays on a xhibit central station (xcs) went blank while monitoring telemetry patients.Spacelabs healthcare technical support assisted the customer in troubleshooting the xcs and the issue was traced down to faulty display cables.Once the display cables were replaced, the customer confirmed the displays were now functioning properly.The failure was due to bad display cables.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XHIBIT CENTRAL STATION
Common Device Name
CENTRAL STATION REMOTE MONITOR
Manufacturer (Section D)
SPACELABS HEALTHCARE
35301 se center st
snoqualmie WA 98065
Manufacturer (Section G)
SPACELABS HEALTHCARE
35301 se center st
snoqualmie WA 98065
Manufacturer Contact
whitney snyder
35301 se center st
snoqualmie, WA 98065
MDR Report Key23854474
Report Number3010157426-2025-00240
Device Sequence Number17416375
Product Code MHX
UDI-Device Identifier10841522100345
UDI-Public(01)10841522100345(11)170826
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K122146
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 12/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number96102
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/04/2025
Initial Report FDA Received Date12/19/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured08/26/2017
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1