It was learned through implant patient registry and investigation that a patient with a 33mm 7300tfx mitral valve was explanted after an implant duration of 9 years, 1 month due to stenosis.The patient presented with shortness of breath, and vertigo.The explanted valve was replaced with a 33mm 11400m valve.Patient was in recovery post procedure.The patient was discharged on pod#9.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional customer complaint.The information reported may or may not be related to the edwards device.
|
|
It was learned through implant patient registry and investigation that a patient with a 33mm 7300tfx mitral valve was explanted after an implant duration of 9 years, 1 month due to calcification, with pannus and stenosis.The patient presented with shortness of breath, vertigo and endurance.The explanted valve was replaced with a 33mm 11400m valve.Patient was in recovery post procedure.The patient was discharged on pod#9.Per pathology report: final diagnosis with calcification and fibrosis.Gross description; carpentier-edwards mitral valve bioprosthesis with calcifications, with no cusp tears, without thrombus, with obstructive fibrous ingrowth (pannus), and without vegetations.
|
|
Updated sections: b5, g3, g6, h6 device code(s), type of investigation, investigation findings, and investigation conclusions.A device history record (dhr) review is not required because there is no allegation of a malfunction which could be related to a manufacturing non-conformance and/or one was not suspected or confirmed through investigation; no labeling non-conformance/deficiency; no device-related infection; and no evidence of a product failure with regard to design, reliability, or use error.Svd is a logical and expected consequence of the chemical, mechanical, and immunological processes associated with bhv implantation.The common endpoint of all these factors is calcification and degradation.Svd is an acquired condition, intrinsic to bioprosthetic valves, characterized by deterioration of the leaflets or supporting structures, leading to thickening, calcification, tearing, or disruption of the prosthetic valve materials.These changes result in valvular dysfunction manifesting as stenosis and/or regurgitation with a consequent drop in hemodynamic efficacy of the valve.Despite the advancements in the heart valve industry, svd commonly begins 7- 8 years after implantation, with a considerably increased rate in patients with pertinent comorbidities and/or an implant duration of greater than 10 years.Prior investigations and extensive literature review have shown that svd is predominantly related to patient factors and implant duration rather than to manufacturing issues.Events related to device design, manufacturing, or use error tend to manifest at an earlier implant duration.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.Based on the information available, the most likely root cause is patient factors, including an implant duration of 7 or more years.Pannus is a non-immune inflammatory reaction of the body to the implanted prosthesis, characterized by proliferation of fibroelastic tissue and collagen, with a starting point in the suture area and subacute or chronic centripetal evolution.Pannus can have both beneficial and harmful effects depending on the amount of growth and location.A small amount of host tissue growth over the suture line is expected and is needed to form a non-thrombogenic surface and complete the healing process after device implantation.In contrast, an excessive amount of pannus growth can cause nonstructural dysfunction that may require intervention.Host tissue growth can extend onto the cusp surfaces leading to thickening of the cusps, cusp retraction or curling, leaflet immobility, and/or abnormal coaptation, potentially resulting in valvular regurgitation, elevated gradients, and/or stenosis.Pannus/host tissue overgrowth is most likely due to patient factors and is unlikely to be related to a manufacturing non-conformance.There is no evidence to suggest a manufacturing defect caused or contributed to the reported event.The most likely root cause is patient factors.
|