| Catalog Number |
363047 |
| Medical Device Problem Code |
Overfill (2404)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® plus citrate tube, an unspecified number of tubes appear to be overfilled.There was no health impact or consequences reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: bd received 3 photos for investigation.Evaluation of the three photos indicated a satisfactory draw level.Sufficient volume for citrate tubes is achieved when the blood draw falls above the minimum fill indicator and below the bottom of the shield.Additionally, 40 retained samples were subjected to functional draw volume testing and all tubes tested drew within specification.Based on a review of the device history record for the incident lot, all product specifications and requirements for lot release were met.This complaint has not been confirmed for the indicated failure mode: overfill.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored.Our business team regularly reviews the collected data for the identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported while using bd vacutainer® plus citrate tube, an unspecified number of tubes appear to be overfilled.There was no health impact or consequences reported.
|
| |
|
Search Alerts/Recalls
|