• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. NAVLOCK; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. NAVLOCK; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 9734683
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/28/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H2 additional information: b5 updated.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
It was reported that all 4 colors had slight scratches on the internal side which could not be captured in an image.All navlocks trackers were difficult to attach or remove, especially the grey one was hard and was most likely stuck as it cannot be removed.
 
Additional Manufacturer Narrative
Select patient information cannot be provided due to regional privacy regulations.G2: foreign country - japan h3, h6: no products have been returned to medtronic for analysis.Codes b17, c20, and d15 are applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
Medtronic received information regarding a navigation system being used during a cervical spine procedure.It was reported that the user inserted the infinity drill bit into the navlock tracker violet and created a pilot hole, but it stopped rotating.Attached it to another navlock tracker, but it became stuck again.There was no surgical delay.There was no impact on patient outcome.
 
Additional Manufacturer Narrative
H3, h6: the tracker was returned for evaluation.Analysis found a mechanical failure.The returned tracker had galling inside the handle that would not allow the insertion of instruments.Otherwise, with markers attached and fully seated, the tracker displayed a good geometry error with normal tracking.Codes b01, c07 and d02 are applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVLOCK
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC.
200 medtronic drive
lafayette CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC.
200 medtronic drive
lafayette CO 80027
Manufacturer Contact
justin ellis
8200 coral sea street ne mvs21
mounds view, MN 55112
7635265677
MDR Report Key23857386
Report Number1723170-2025-03952
Device Sequence Number5030727
Product Code OLO
UDI-Device Identifier00763000054298
UDI-Public(01)00763000054298(11)200821(10)200719
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K171267
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/22/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number9734683
Device Catalogue Number9734683
Device Lot Number200719
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/28/2025
Supplement Date Received by Manufacturer02/12/2026
04/17/2026
Initial Report FDA Received Date12/19/2025
Supplement Report FDA Received Date02/12/2026
04/22/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-