The device was returned to stryker sustainability solutions for evaluation.Upon inspection of the received complaint device, the device handle was intact and showed no signs of visual anomalies.The drive cable housing assembly was intact.No planarity damage to the outer shaft was observed.The returned complaint device was received with the cutting blade in the closed position.The cutting blade functionality was tested manually by twisting the drive cable coil on the drive cable plug to open and close the blade, the blade opened and closed as intended.Using a dvm, the returned complaint device was tested for continuity.With the blade in the closed position, the dvm read shorted (closed) and with the blade open, the device read open.The device was then connected to the myosure controller unit to verify that the device blade would actuate when activated.The device was recognized by the controller and when activated the blade did not actuate.Instead, you could hear the blade/gears spinning the handle.The handle was opened to find the electrical wire kinked at the proximal end of the handle.Per the reported event of, ".¿glitter¿/metal shavings swirling around on the camera and that it was from the myosure device," the blade was inspected with a high magnification camera.During inspection the blade, the blade was found to be sharp.The inner shaft showed no galling marks indicative of metal-on-metal friction.The blade was able to fluctuate up and down within the shaft.A q-tip was used to try and pull any metal shavings from the inner shaft.None were found.The device inspection indicated that the complaint was confirmed for not actuating.A review of the dhr supports that the device met all inspection and test criteria prior to release from stryker.The reported event could be attributed to: mechanical/electrical failure of the device gear assembly or myosure control unit or drive cable connection.Mishandling of device, excessive leverage applied to device.The instructions for use (ifu) state: before using the myosure tissue removal system for the first time, please review all available product information.Use only the myosure control unit to connect to the reprocessed myosure tissue removal device.Use of any other drive mechanism may result in failure of the device to operate or lead to patient or physician injury.If visualization is lost at any point during a procedure, stop cutting immediately.Periodic irrigation of the tissue removal device tip is recommended to provide adequate cooling and to prevent accumulation of excised materials in the surgical site.Ensure that vacuum pressure >200 mm hg is available before commencing surgery.Before using the myosure tissue removal system, you should be experienced in hysteroscopic surgery with powered instruments.Healthy uterine tissue can be injured by improper use of the tissue removal device.Use every available means to avoid such injury.Do not use the reprocessed myosure tissue removal device to resect tissue that is adjacent to an implant.When resecting tissue in patients that have implants, assure that: the reprocessed myosure tissue removal device¿s cutting window is facing away from (i.E.180° opposite) the implant.The visual field is clear; and the reprocessed myosure tissue removal device¿s cutting window is engaged in tissue and is moved away from the implant as tissue resection proceeds.In the event an implant becomes entangled with a myosure cutter, the following steps are recommended: cease cutting immediately: kink the reprocessed myosure tissue removal device¿s outflow tube to prevent a loss of uterine distension.Disconnect the reprocessed myosure tissue removal device¿s drive cable from the control box.Grasp the end of the reprocessed myosure tissue removal device drive cable with a hemostat or other clamping device.Hold the drive cable hub and tissue removal device to prevent twisting.Open the tissue removal device¿s cutting window by manually twisting the hemostat counterclockwise; and gently pull the reprocessed myosure tissue removal device into the hysteroscope to detach the myosure device from the implant.Operation: 1.Push the power switch to the on position.2.The foot pedal activates tissue removal device operation.The foot pedal turns the motor on and off.Once the foot pedal is depressed, the reprocessed tissue removal device accelerates and rotates to the set speed and continues until the foot pedal is released.3.Press the foot pedal and observe the reprocessed tissue removal device action to verify that the motor runs and that the cutting window is closed.4.Introduce the reprocessed tissue removal device through the straight 3 mm working channel of a hysteroscope.5.Under direct hysteroscopic visualization, position the reprocessed tissue removal device¿s side facing cutting window against target pathology.6.Press the foot pedal to activate the reprocessed tissue removal device¿s cutting blade.7.The reprocessed tissue removal device¿s reciprocating action alternately opens and closes the device¿s cutting window to the vacuum flow thereby drawing tissue into the cutting window.8.Cutting takes place when the reprocessed tissue removal device cutting edge rotates and translates across the reprocessed tissue removal device¿s cutting window.Caution: excessive leverage on the reprocessed tissue removal device does not improve cutting performance and, in extreme cases, may result in wear, degradation, and seizing of the cutter assembly.The reported event will continue to be monitored through post-market surveillance.
|
|
It was reported that the myosure operated as normal at the beginning of the case.The patient was high in the bed and the doctor had a hard time visualizing, so they had to stop the procedure and move the patient down.Upon resuming the end of the procedure, the staff and the doctor said there was ¿glitter¿/metal shavings swirling around on the camera and that it was from the myosure device." the circulator stated they believed the metal shavings were flushed out from the fluid management system.It could not be confirmed since the shavings were so small.No other action was taken by the surgeon or the team for the metal shavings.There was no patient injury, medical intervention, or extended procedure time reported.These are commonly used devices that are readily available.
|