| Catalog Number |
468-350 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
10/08/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: deflation.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported deflation.Later, healthcare professional reported "textured device".This record is for the right side.The device has been explanted and replaced.The device will be returned.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported deflation.Later, healthcare professional reported "textured device".This record is for the right side.The device has been explanted and replaced.The device will be returned.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the device related to the reported events anxiety-product/procedure and deflation was received on january 26, 2026 with lot number 1651541.Per the investigation procedure, the device is analyzed through visual inspection microscopic inspection if openings are observed and a weight verification.Per the analysis performed, the assessments of the complaints and any potential manufacturing issue are displayed along with any further actions required: ¿ anxiety-product/procedure: unable to observe through visual inspection as it is a physiological phenomenon.¿ deflation: observed an opening through microscopic inspection assessed as fold flaw opening.No further actions are required as it is not related to the manufacturing process.None of the other observations performed during the device analysis creases are found to be potentially related to the manufacturing process, and, therefore, no further actions are required for these observations.
|
| |
|
Search Alerts/Recalls
|