| Model Number |
N161 |
| Medical Device Problem Codes |
Delayed Charge Time (2586); Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/18/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d), with 8 months of remaining longevity, had reached the elective replacement indicator (eri) and is now at end of life (eol).Technical services (ts) reviewed the case and confirmed that eri was triggered due to charge time (ct).Ts also stated that the patient likely experienced a connection problem; otherwise, the alert would have been sent via latitude.This device remains in service.No adverse patient effects were reported.Additional information was received stating this device was explanted and replaced.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This product was explanted and has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.It was confirmed that this device met manufacturing specifications prior to distribution and there was no indication that the device manufacturing process contributed to the reported complaint.A review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.A risk review was completed and confirmed that the event of delayed charge time was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance to labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of device problem excluded.If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that this cardiac resynchronization therapy defibrillator (crt-d), with 8 months of remaining longevity, had reached the elective replacement indicator (eri) and is now at end of life (eol).Technical services (ts) reviewed the case and confirmed that eri was triggered due to charge time (ct).Ts also stated that the patient likely experienced a connection problem; otherwise, the alert would have been sent via latitude.This device remains in service.No adverse patient effects were reported.Additional information was received stating this device was explanted and replaced.No additional adverse patient effects were reported.
|
| |
|