It was reported that an implant was removed due to a sinus perforation and sinus infection.Abscess, pain, edema and inflammation were reported as a result of the event.Sinus lift was performed, sinus had infection and the implant was removed.
|
|
Zimvie complaint number (b)(6).The following fields have been updated: b4: date of this report.B5: describe event or problem.G3: date received by manufacturer.G6: type of report.H1: type of reportable event.H2: follow up type.H3: device evaluated by manufacturer.H6: adverse event problem.H11: additional narrative.Zimvie received one (1) tsx54b10, (tsx implant, 5.4mmd, 10mml) for evaluation.Visual evaluation / functional test was performed; the implant has minor signs of use.There was no damage identified or signs of malfunction that would have contributed to the event.Measurements match drawing.Dhr review was completed for the subject lot number 2120026524.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2120026524 for similar events and no other complaint was identified.Review completed utilizing keywords: ¿dental: medical: perforated sinus and dental: medical: infection¿ based on the investigation and risk management file review, the most likely root cause determined from the investigation was customer error and/or patient factors.For infection: a summary investigation has been completed for infection events identifying that a definitive root cause cannot be identified due to the wide range of external (non-design/ non-manufacturing related) factors potentially impacting the sterility of the implant and its environment.Therefore, based on the available information, a device malfunction did not occur.No damage to the implant was identified.The reported event was non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
|