Catalog Number 8065752134
Medical Device Problem Code Suction Problem (2170)
Health Effect - Clinical Code Insufficient Information (4580)
Type of Reportable Event Malfunction
Event or Problem Description
A physician reported that no aspiration of probe was performed during the cataract [with intraocular lens (iol) implant] surgery.The surgery was completed on the same day after replaced with the same product.There was no report of patient impact.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.3., h.6.And h.11.No technical inquiries were received from the customer.One opened probe, with tip protector, in a tray with other items was received for the report.The sample was visually inspected and found to be nonconforming with orange/brown foreign material on the port face and inside the port.The sample was then functionally tested for actuation and aspiration.The sample was found to be conforming for actuation and nonconforming for aspiration.The probe was disassembled and the components inspected.Nominal wear observed on inner cutter when compared to the degree of wear based on continuous actuation of the probe visual standard photos.Gouge marks observed at several area.The sample was tested with a syringe and resistance was felt.A wire was inserted through the aspiration path.The sample was retested with a syringe, and resistance was felt.Solid material was observed to be blocking the aspiration path and cannot be removed for further evaluation.A review of the device history record traceable to the lot number obtained from the device¿s rfid (radio frequency identification) tag, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified, and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.The complaint evaluation confirms the probe had an aspiration failure.The root cause of the aspiration failure is due to foreign material in the aspiration path.How and when foreign material was introduced into the aspiration path cannot be determined from this evaluation; however, the foreign material is most likely surgical material from the surgical use of the probe.The sample met specification for actuation and cut and the root cause and its origin are inconclusive for the aspiration issue, further investigative or manufacturing actions are not warranted at this time.All probes are 100 percentage visually inspected and tested for actuation, aspiration, and cut during manufacturing.Any non-conformances found are removed from the lot and scrapped.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION ULTRAVIT PROBE PROCEDURE PAK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23861862
Report Number1644019-2025-05612
Device Sequence Number10006699
Product Code LRO
UDI-Device Identifier00380657521340
UDI-Public(01)00380657521340(17)260228(11)240308(10)16LJ4F
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2026
Device Catalogue Number8065752134
Device Lot Number16LJ4F
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/25/2025
Supplement Date Received by Manufacturer04/01/2026
Initial Report FDA Received Date12/22/2025
Supplement Report FDA Received Date04/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/08/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
CENTURION VISION SYSTEM.
Patient SexUnknown
-
-