| Model Number |
ONYXNG25018UX |
| Medical Device Problem Codes |
Difficult to Advance (2920); Material Deformation (2976); Positioning Problem (3009)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a procedure involving one onyx frontier coronary drug eluting stent, stent deformation occurred in vivo during positioning/advancement.The stent was removed from the hoop and advanced through the guide to the proximal rca, where it was unable to cross the lesion.The stent was removed from the body and flushed and re advanced when it was noticed that the stent surface was uneven.The lesion was located in the proximal right coronary artery (rca) with moderate tortuosity and moderate calcification.The device was inspected prior to use with issues noted.The lesion was pre-dilated.The device did not pass through a previously deployed stent.Resistance was encountered when advancing the device and excessive force was not used during delivery.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis: the device was returned for evaluation.The stent was positioned on the balloon between the marker bands in accordance with the positioning specification.However, due to deformation, the stent did not meet the visual acceptance criteria.Deformation was observed on the mid stent wraps, with struts raised.The inner lumen patency was verified with a mandrel.No deformation was evident on the distal tip.No other damage was observed on the remainder of the device.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|