• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES PRODUCTS LLC ADULT CRANIOTOME; MOTOR, DRILL, ELECTRIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES PRODUCTS LLC ADULT CRANIOTOME; MOTOR, DRILL, ELECTRIC Back to Search Results
Catalog Number CRANI-A-G1
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11 additional narrative: the actual device has been returned and is currently pending evaluation.Once the evaluation has been completed, a supplemental medwatch report will be sent accordingly.
 
Event or Problem Description
It was reported that the motor device overheated.It was not reported if the device was used in surgery.It was not reported if there were any delays in a surgical procedure, or if a spare device was available for use.The exact date of this event was unknown but was noted to have occurred in 2025.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: a visual and functional assessment was performed: the following steps failed: visual assessment.Cutter insertion.During service/repair the following was observed: rigid guard scratched; color label scratched; bar cannot be inserted (bearing detached).During the service evaluation the following failures were identified: visual : craniotome neuro-tip bent/damaged.Visual : color band chipping/fading.Visual : bearing damaged.Device interaction (2+ devices) : cannot insert cutter.Conclusion: ¿ failure reported is not confirmed.¿ root cause: component wear.¿ action: exchange, replacement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADULT CRANIOTOME
Common Device Name
MOTOR, DRILL, ELECTRIC
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
elba bello
4500 riverside drive
palm beach gardens, FL 33410
9083863534
MDR Report Key23862330
Report Number1045834-2025-00650
Device Sequence Number9387001
Product Code HBC
UDI-Device Identifier00845384016410
UDI-Public(01)00845384016410
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K133604
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/19/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCRANI-A-G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/12/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/12/2025
Supplement Date Received by Manufacturer12/26/2025
Initial Report FDA Received Date12/22/2025
Supplement Report FDA Received Date12/26/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/04/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-