| Model Number |
25131-5 |
| Medical Device Problem Code |
Complete Blockage (1094)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/20/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a valve.It was reported that the valve was occluded at the time of occlusion testing.When injecting to check whether the valve system was patent, the physician felt significant resistance during injection.After connecting the catheter, no fluid flowed despite very high pressure.
|
| |
|
Additional Manufacturer Narrative
|
|
H3.Product analysis determined that the visual and functional inspection confirmed the returned device was patent.The valve passed requirements for patency and leak testing.No signs of damage or debris were observed on the valve.The valve was not able replicate the failure condition of no flow as observed in the field.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|