| Catalog Number |
ILPC344U |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/30/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that fracture screw got broken.Tooth site # 35.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number: (b)(4).A2: age and date of birth unknown / not provided.A4: patient weight unknown / not provided.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie received one (1) ilpc344u, (certain low profile one-piece abutment 3.4mm(d) x 4mm(h)) for evaluation.Visual evaluation was performed, a fracture has been identified at the threads.Dhr review was completed for the subject lot number 2023050601.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2023050601 for similar events and no other complaint was identified.Review completed utilizing keywords: dental : functional : fracture : screw the customer did not submit images for the reported event.Based on the investigation and risk management file review the most likely root cause determined from the investigation was clinician incorrectly engages abutment screw into implant.Therefore, based on the available information, a device malfunction did occur.The fracture has been identified on the threads.The reported event is confirmed.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.The following sections have been updated: b4: date of this report.D9: device availability.G3: date received by manufacturer.G6: checked "follow-up".H2: checked follow-up type.H3: changed "no" to "yes".H6: entered evaluation codes.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie did not receive one (1) ilpc344u, (certain low profile one-piece abutment 3.4mm(d) x 4mm(h)) for evaluation.Visual inspection could not be performed.The investigation has been performed based on the available information using the item number, device history record (dhr) review, and risk management file (rmf).Dhr review was completed for the subject lot number 2023050601.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2023050601 for similar events and no other complaint was identified.The customer did not submit images for the reported event.Based on the investigation and risk management file review the most likely root cause determined from the investigation was clinician incorrectly engages abutment screw into implant.Therefore, based on the available information, a device malfunction could not be verified.Without device receipt, the reported event is non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.The following sections have been updated: b4: date of this report.G3: date received by manufacturer.G6: checked "follow-up".H6: entered evaluation codes.H10: added manufacturer narrative.
|
| |
|
Event or Problem Description
|
|
No further event information available at the time of this report.
|
| |
|
Search Alerts/Recalls
|