| Catalog Number |
136551000 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Joint Dislocation (2374)
|
| Date of Event |
12/04/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that patient hip was dislocated.Upon view of the patient x-ray, it was found that the acetabular component was malpositioned.The cup was vertical and dr.Wanted to revise the acetabular component.When asked if he had any information on the patients initial surgery, he said he didn¿t have any of the information.He was unsure of where the patient had their surgery, who did it, or when it was done.The cup was removed and the patient was prepared and implanted with a revision multihole cup.An elevated rim liner was implanted and a trial reduction was performed.Stability was confirmed, so the real ts ceramic head was opened and implanted.Hip was reduced and the closing process began.No delays or adverse events took place.Doi: unknown.Dor: unknown.Affected side: right hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary- no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required product/lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot- the device lot number is unknown, therefore a device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.Added: b1 (product problem), d10 (concomitant).
|
| |
|
Search Alerts/Recalls
|
|