SMITH & NEPHEW, INC. Q-FIX KNTLS ASA 1.8MM W/1 MT SUT BL; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
| Catalog Number |
72205884 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Perforation (2001)
|
| Date of Event |
11/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, during an arthroscopic acl reconstruction and lateral extra-articular tenodesis, after preparing the insertion site with a 1.8 mm q-fix drill kit, the surgeon pulled three sutures together after two (2) q-fix knotless anchors were deployed, and the anchors were still inside the drill hole.However, when shortening the blue minitape to secure the tendon, the whole anchors were out from the femur drill hole.The procedure was resumed, after a non-significant delay, using an equivalent s+n back-up, in an additionally drilled bone hole.No void was left inside the patient.No further complications were reported.The patient has been discharged.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference number: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a patient complication related to a product problem with a smith+nephew device.The reported issue(s) relate to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Search Alerts/Recalls
|
|
|