• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF SINGLEPIECE IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF SINGLEPIECE IOL; INTRAOCULAR LENS Back to Search Results
Model Number SA60AT
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Capsular Bag Tear (2639)
Date of Event 11/26/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A product was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non-health care professional reported a capsule rupture in the patient.The physician removed the lens and replaced with a three-piece company lens.Additional information received stating that there was no patient harm.
 
Additional Manufacturer Narrative
Additional information was provided in d10., h.3., h.6.And h.11.The product was not returned.A device history record review and a non-conformance review of the reported lot number was conducted.No deviations were identified and all corresponding production release specifications defined in the device master record were met.The file indicates the use of a qualified cartridge and handpiece with a non-qualified viscoelastic.We are unable to verify the reported issue of a capsule rupture.The product was not returned for evaluation.The file indicates the use of a non-qualified viscoelastic.The company foldable intraocular lens (iol) are qualified for use with a company qualified delivery system (handpiece and cartridge) and ophthalmic viscosurgical device (ovd) combination.The instructions for use (ifu) instructs that a company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The ifu instructs: using holding forceps, grasp the lens by the optic edge and gently place the lens anterior side up into the back of the ovd-filled cartridge.The lens should be inserted until the optic is a little more than half-way inside the cartridge.Use the holding forceps to gently push down on the lens, verifying that the lens is on the bottom surface of the cartridge.Using holding forceps, take the trailing haptic, and gently fold the haptic onto the anterior side of the optic.Slowly grip or push the optic edge to position the lens as far into the cartridge as the forceps will permit, while ensuring the lens remains on the bottom surface of the cartridge and the trailing haptic remains on the optic.Failure to follow these steps may cause the lens to advance incorrectly causing delivery issues and/or damage.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF SINGLEPIECE IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23864857
Report Number1119421-2025-03209
Device Sequence Number4437960
Product Code HQL
UDI-Device Identifier00380652555814
UDI-Public(01)00380652555814(17)300723(11)250725(10)16074824
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Reporter Type Manufacturer
Report Source Foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/06/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberSA60AT
Device Catalogue NumberSA60AT.205
Device Lot Number16074824
Was Device Available for Evaluation? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/26/2025
Supplement Date Received by Manufacturer01/13/2026
Initial Report FDA Received Date12/22/2025
Supplement Report FDA Received Date02/06/2026
Date Device Manufactured07/25/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ACRYSOF MULTIPIECE IOL.; AMVISC PLUS 0.8ML.; MONARCH II IOL DELIVERY SYSTEM, INJECTOR.; MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE C.
Outcome Attributed to Adverse Event Other;
Patient Age46 YR
Patient SexUnknown
-
-