(b)(4).D10: stem implant, #item 00598801013, #lot 62370149.Unk femoral component, #item unknown, #lot unknown.Unk tibial component, item #unknown, #lot unknown.Unk fixation screw, #item unknown, #lot unknown.Unk tibial spacer, #item unknown, #lot unknown.Unk femoral spacers, #item unknown, #lot unknown.Unk insert, #item unknown, #lot unknown.Therapy date: (b)(6) 2025.G2: foreign source country: italy.H6: component code: mechanical (g04) - stem.An investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4,b5,d2,d9,g1,g3,g6,h1,h2,h3,h6
the following sections were corrected:d1,d10
d10:
stem, #item 00598801013, #lot 62370149
unk femoral component, #item unknown, #lot unknown
unk tibial component, item #unknown, #lot unknown
unk fixation screw, #item unknown, #lot unknown
unk tibial augment, #item unknown, #lot unknown
unk femoral augment, #item unknown, #lot unknown
unk insert, #item unknown, #lot unknown
therapy date: (b)(6) 2025
the cmp was confirmed
offset stem pn 00598802011: visual examination of the returned product identified signs of use.Products were received assembled.Unable to verify taper or offset while assembled.Part visually matches the blueprint.Dimensional analysis of the product determined that the product, where measured, was conforming to print specifications.Stems: review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Insufficient information provided.Unable to perform a compatibility check.Offset stem pn 00598802011: review of complaint history identified additional similar complaints for the reported item(s) and no additional complaints for the reported part and lot combination(s).Complaints are monitored through monthly complaint review in order to identify potential adverse trends.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following:
imaging findings - ap and lateral views of the right knee show postoperative changes of knee arthroplasty with constrained longstem femoral and tibial components (likely revision tka).The patella is thin and mild patella alta is noted.There is radiolucency around the distal aspect of the tibial stem with lateral positioning of the stem in the tibial diaphysis, suspicious for loosening.Marked osseous demineralization is noted.Impression - revision right knee arthroplasty with radiolucency around the tibial stem and varus alignment of the tibial component consistent with loosening.Of note, marked radiographic osteopenia is noted, which can predispose to implant loosening.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|