| Model Number |
M0068318261 |
| Medical Device Problem Code |
Misfire (2532)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/24/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Block h6: device code a050502 captures the reportable event of device misfire.
|
| |
|
Event or Problem Description
|
|
It was reported that during a sacrospinous ligament fixation procedure using a capio slim device, the device misfired.The procedure was completed with another of the same device.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that during a sacrospinous ligament fixation procedure using a capio slim device, the device misfired.The procedure was completed with another of the same device.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Block h11: upon receipt at our quality assurance laboratory, this capio device underwent a thorough analysis.Visual inspection of the device did not have any visual damage/defect.During functional inspection, the carrier was able to move in and out; after deploying the carrier, the cage was able to catch the suture.Based on the information available and analysis results, the reported allegations of device misfire could not be confirmed through product analysis.A conclusion code of no problem detected was assigned to this investigation since after visual and functional inspections in the complaint device it was not possible to identify any evidence of either the alleged issue or any defect on the device.Block h6: device code a050502 captures the reportable event of device misfire.
|
| |
|