| Model Number |
9-ASD-019 |
| Medical Device Problem Code |
Difficult to Fold, Unfold or Collapse (1254)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2025, a 19mm amplatzer septal occluder was chosen for implant using a 8f amplatzer trevisio delivery system.During the procedure, while implanting, it was observed that the device conformed into a cobra deformation.The deformed device was released from the delivery cable inside the patient.There was no interaction with the cardiac structures during deployment and no angulation/kink were noticed in the delivery system.The procedure was completed using a new 20mm amplatzer septal occluder.The patient remained hemodynamically stable throughout the procedure.There was no clinically significant delay in procedure and no adverse patient effects.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of device deformity was reported.A returned device assessment could not be performed as the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all defined manufacturing specifications.There were no complaints associated with any other devices from the lot.Based on the information received, the cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Search Alerts/Recalls
|