Model Number 408718
Medical Device Problem Code Failure to Capture (1081)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/12/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Combination product: yes.Upon receipt, the lead under complaint was found cut 6.5 cm distal to the df4 connector pin.The proximal fragment was received for analysis.The distal fragment was not received for analysis.The returned lead fragment was analyzed.The inspection revealed that the insulation was, apart from the present cut, free of breaches.Further analysis of the returned lead fragment did not reveal any irregularity that might have contributed to the reported clinical observations.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were re-investigated.All production steps were performed accordingly, and in particular the final acceptance test proved the device functions to be as specified.In conclusion, the analysis did not reveal any sign of a material or manufacturing problem.
 
Event or Problem Description
This lv lead was explanted due to failure to capture in any vector.No adverse patient events were reported.Should additional information be received, this file will be updated.
 
Additional Manufacturer Narrative
Combination product: yes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENTUS PROMRI OTW QP L-75/49
Common Device Name
LV LEAD
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key23873534
Report Number1028232-2025-06258
Device Sequence Number4438272
Product Code OJX
UDI-Device Identifier04035479148942
UDI-Public(01)04035479148942(17)220228
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P070008
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2022
Device Model Number408718
Device Catalogue Number408718
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/24/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/09/2025
Supplement Date Received by Manufacturer02/13/2026
Initial Report FDA Received Date12/22/2025
Supplement Report FDA Received Date02/16/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/21/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age74 YR
Patient SexFemale
-
-