• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ IMAGING CCU, MODULAR, IMAGE1 S CONNECT II, US

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ IMAGING CCU, MODULAR, IMAGE1 S CONNECT II, US Back to Search Results
Model Number TC201US
Medical Device Problem Codes Circuit Failure (1089); Insufficient Device Problem Information (3190)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/03/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device will be forwarded to the manufacturing site for investigation.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted unsolicited.This event is filed under internal complaint id (b)(4).Cross reference complaint ids: (b)(4).
 
Event or Problem Description
It was reported that "the system froze up completely during a case yesterday, in which they had to reboot the system".It was confirmed that there was no patient impact or injury and the surgeon was able to complete the procedure.No negative impact in state of health reported.Cross reference mdrs: 2027009-2025-02399.2027009-2025-02400.
 
Additional Manufacturer Narrative
Device evaluation: the complaint issue was described as per se - on our very first day of cases on (b)(6) 2025 we received an error message that states menu temporarily unavailable cycle power.The customer's complaint was verified.The customer's system was verified to produce the "menu temporarily unavailable" error message.The customer returned the following devices which were tested together: tc201us, tc301us, and tc304us.The "menu temporarily unavailable" error message was displayed during two occasions.The mtu message occurs when the tc201us davinci processor stops responding.There is no known component level rework to address the mtu issue, so a tc201us motherboard replacement is required to correct the complaint.The root cause was determined as tc201us has intermittent motherboard causing menu temporarily unavailable message.This event is filed under internal complaint id (b)(4).Cross reference complaint ids: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CCU, MODULAR, IMAGE1 S CONNECT II, US
Common Device Name
CCU, MODULAR, IMAGE1 S CONNECT II, US
Manufacturer (Section D)
KARL STORZ IMAGING
1 s. los cameros road
goleta CA 93117
Manufacturer (Section G)
KARL STORZ IMAGING
1 s. los cameros road
goleta CA 93117
Manufacturer Contact
anja fair
2151 e grand ave
el segundo, CA 90245
4242188247
MDR Report Key23874033
Report Number2027009-2025-02398
Device Sequence Number5571087
Product Code FET
UDI-Device Identifier04048551422625
UDI-Public04048551422625
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K233333
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberTC201US
Device Catalogue NumberTC201US
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/03/2025
Supplement Date Received by Manufacturer03/02/2026
Initial Report FDA Received Date12/22/2025
Supplement Report FDA Received Date03/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-