W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Catalog Number |
CXT361414 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Codes |
Aneurysm expansion (4597); Clinical Endoleak (4843)
|
| Date of Event |
11/05/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A1: no patient specific details have been provided.Therefore, the patient initials reflect the w.L.Gore internal case number.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
The following was reported to gore on december 8, 2025, from the imedidata database: on (b)(6) 2024, this 77-year-old male patient underwent endovascular treatment for an abdominal aortic aneurysm.The patient was treated with gore® excluder® conformable aaa endoprosthesis and gore® excluder® aaa endoprosthesis devices.The gore® devices were successfully navigated to the intended location and deployed.On (b)(6) 2024, the patient developed a type ii endoleak.On (b)(6) 2025, a ct revealed that the maximum diameter of the abdominal aorta was 70 mm.On (b)(6) 2025, a duplex ultrasound confirmed the type ii endoleak.On (b)(6) 2025, a ct revealed that the maximum diameter of the abdominal aorta was 75 mm.On (b)(6) 2025, a cta revealed that the type ii endileak and the maximum diameter of the abdominal aorta was 75 mm.On (b)(6) 2025, a reintervention was performed to treat the type ii endoleak.Coil embolization was performed, and the endoleak was successfully resolved.
|
| |
|
Additional Manufacturer Narrative
|
|
Emdr section h6, code c19 updated to reflect results of product history review.H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.
|
| |
|