Catalog Number CEB231210A
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/01/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
On (b)(6) 2025, this patient underwent endovascular treatment for an abdominal aortic aneurysm using gore® excluder® aaa endoprosthesis and gore® excluder® iliac branch endoprosthesis.While deploying the contralateral leg component by pushing it up in the ipsilateral side, it moved proximally and was placed where the proximal side of the device was extended approximately 1cm past the flow splitter.There was no obstruction of blood flow in the contralateral side; however, a bare stent was placed for preventive purposes from the same level.After the iliac branch component(ibc) was fully deployed, ballooning was performed using the kissing ballooning technique(kbt).Reportedly, the ipsilateral leg was compressed by the contralateral leg, and the lumen of the ipsilateral leg became narrowed.Attempts were made to expand the ipsilateral leg using kbt, but it was unsuccessful.Therefore, a bare stent was placed.The patient tolerated the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® EXCLUDER® ILIAC BRANCH ENDOPROSTHESIS
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL SILICON VALLEY B/P
2890 de la cruz blvd.
santa clara CA 95050
Manufacturer Contact
hiroko sakamiya
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key23874167
Report Number3013164176-2025-02842
Device Sequence Number21450042
Product Code MIH
UDI-Device Identifier00733132635290
UDI-Public(01)00733132635290(17)280721
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCEB231210A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/01/2025
Initial Report FDA Received Date12/22/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/23/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age88 YR
Patient SexMale
-
-