W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® ILIAC BRANCH ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Catalog Number |
CEB231210A |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H6: a review of the manufacturing records for the device verified that the lot met all pre-release specifications.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2025, this patient underwent endovascular treatment for an abdominal aortic aneurysm using gore® excluder® aaa endoprosthesis and gore® excluder® iliac branch endoprosthesis.While deploying the contralateral leg component by pushing it up in the ipsilateral side, it moved proximally and was placed where the proximal side of the device was extended approximately 1cm past the flow splitter.There was no obstruction of blood flow in the contralateral side; however, a bare stent was placed for preventive purposes from the same level.After the iliac branch component(ibc) was fully deployed, ballooning was performed using the kissing ballooning technique(kbt).Reportedly, the ipsilateral leg was compressed by the contralateral leg, and the lumen of the ipsilateral leg became narrowed.Attempts were made to expand the ipsilateral leg using kbt, but it was unsuccessful.Therefore, a bare stent was placed.The patient tolerated the procedure.
|
| |
|
Search Alerts/Recalls
|
|
|