| Catalog Number |
8065752180 |
| Medical Device Problem Code |
Suction Failure (4039)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
11/24/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is:(b)(4).
|
| |
|
Event or Problem Description
|
|
A health care professional reported that there was no vacuum during cataract surgery (with intra ocular lens implantation).New cassettes were required to proceed with the surgery.There was no information about patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided in d.9., h.3., h.6.And h.11.Two unused fluidics management system in the trays were visually inspected, and no obvious defects were observed.The drain bag was securely adhered on the cover.Both irrigation and aspiration luers were intact.Flare shape was observed in the blue socket fitting at the aspiration tubing insertion end.The products were tested using the console with the current software procedure.The products could be recognized by the console.The products primed and tuned with the handpiece and the 0.9 milli meter aspiration bypass system tip and infusion sleeve from the lab stock successfully and the service data could be retrieved.No system message code displayed on console screen.Cassette maximum vacuum and aspiration flow rate, and irrigation flow rate were measured and met specifications.No problem was found with the returned cassette fluidics.The product functioned within specification per procedure the investigation conducted a non-conformance review of the reported lot number.No deviations that could have contributed to this event were identified and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information and materials received; the investigation was unable to identify the root cause or origin of the reported event as the product functioned per specifications.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.No further investigative or manufacturing actions are warranted at this time due to the product meeting specifications.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional new information was received that the event occurred during the phaco emulsification mode of the cataract surgery (b.5.Updated).The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
Additional new information was received that the event occurred during the phaco emulsification mode of the cataract surgery.
|
| |
|