W. L. GORE & ASSOCIATES, INC. GORE® EXCLUDER® AAA ENDOPROSTHESIS; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
|
Back to Search Results |
|
| Catalog Number |
PLC201200 |
| Medical Device Problem Codes |
Device Dislodged or Dislocated (2923); Device Endoleak (4074)
|
| Health Effect - Clinical Codes |
Obstruction/Occlusion (2422); Clinical Endoleak (4843)
|
| Date of Event |
12/16/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A1: no patient specific details have been provided.Therefore, the patient initials reflect the w.L.Gore internal case number.D10: concomitant medical products: (relevant associated devices used with the device being reported on): gore® excluder® conformable aaa endoprosthesis (main trunk component).Product history review and imaging evaluation are ongoing.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Event or Problem Description
|
|
It was reported to gore that on (b)(6) 2025, a patient with curved anatomy underwent an endovascular aortic aneurysm repair (evar) using a gore® excluder® conformable aaa endoprosthesis (main trunk component) and a gore® excluder® aaa endoprosthesis (contralateral leg component).The devices were successfully implanted.On (b)(6) 2025, it was reported that the gore® excluder® aaa endoprosthesis (contralateral leg component) was disconnected from the main trunk, without any migration of the gore® excluder® conformable aaa endoprosthesis (main trunk component).A reintervention was performed on (b)(6) 2025.A device of unknown brand was used to reline the gore® excluder® aaa endoprosthesis - contralateral leg component) and implanted as a bridge stent between it and the main trunk component.The procedure was successfully concluded.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated h6: added health effect - clinical code e2328.Added medical device problem code a051201.Updated investigation findings code to c24, code c21 is no longer applicable.Updated investigation conclusions code to d15, code d16 is no longer applicable.As part of our routine quality procedures, each batch of devices undergoes comprehensive quality control testing and inspections prior to release for distribution.Gore reviewed the manufacturing records associated with the reported lot number and verified that all release criteria had been met.The primary reported device failure mode, disconnection of the contralateral leg endoprosthesis from the trunk-ipsilateral leg endoprosthesis within one week of implant, is consistent with the evaluation of the provided clinical images.The clinical images only represent one time point five days after device implant so device migration cannot be fully assessed, but the contralateral leg is disconnected 5 cm from the trunk-ipsilateral leg.The contralateral leg is also observed in the clinical imaging to be occluded.The cause for the separation of the contralateral leg endoprosthesis from the conformable excluder trunk-ipsilateral leg could not be established with the available information.
|
| |
|