• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES GMBH LOCKSCR Ø5 L38 F/NAILS TAN LIGHT GREEN; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES GMBH LOCKSCR Ø5 L38 F/NAILS TAN LIGHT GREEN; SCREW, FIXATION, BONE Back to Search Results
Catalog Number 04.005.528
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Codes Unspecified Infection (1930); Unspecified Tissue Injury (4559)
Date of Event 11/29/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported a revision surgery occurred on (b)(6) 2025 due to osteomyelitis.Patient also presented with exposure of the screws in the ankle causing soft tissue injury.The patient remains under management for irrigation and evaluation for possible amputation depending on the evolution of their vascular condition.Date of implant: (b)(6) 2024.Date of explant: (b)(6) 2025.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: if information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H11 additional narrative: h3, h6: photo investigation: the product was not returned to depuy synthes, however photos were provided for review.The photo and x-ray investigation revealed a chronological history of the patient, screw in implant and revision date did not presented migration.A screw that was not part of the system was observed as well.The e-ifu trauma instructions for use - tfna states the following: combination of medical devices: synthes has not tested compatibility with devices provided by other manufacturers.As the device was not returned, and as-received condition could not be assessed, a dimensional inspection and document/specification review were not completed.The overall complaint was not confirmed as the observed condition of the lockscr ø5 l38 f/nails tan light green would have not contributed to the complained device issue.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review (dhr): lot # provided is not a valid number, therefore, the dhr could not be completed.If device is returned or lot number can be confirmed, the dhr will be revisited.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKSCR Ø5 L38 F/NAILS TAN LIGHT GREEN
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
elba bello
1302 wright lane east
west chester, PA 19380
9083863534
MDR Report Key23875882
Report Number8030965-2025-12867
Device Sequence Number11954947
Product Code HWC
UDI-Device Identifier10886982085962
UDI-Public(01)10886982085962(10)L158
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K172157
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number04.005.528
Device Lot NumberL158
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/03/2025
Supplement Date Received by Manufacturer12/24/2025
Initial Report FDA Received Date12/23/2025
Supplement Report FDA Received Date12/26/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-