| Catalog Number |
AS-IFS1 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Air Embolism (1697); Stroke/CVA (1770)
|
| Date of Event |
11/06/2025
|
|
Type of Reportable Event
|
Death
|
|
Event or Problem Description
|
|
The sales representative reported on behalf of the customer that the device, as-ifs1, airseal ifs, 120v, was being used during a right thoracic case on (b)(6) 2025 when it was reported, ¿embolism caused a stroke¿resulted in a patient death.¿.Further assessment was requested from the reporter, and it was found that the surgery was 2 hours and 26 minutes.Bottle gas was used, and the device pressure was set at 5.No history of sickle cell anemia, and there is no mention of pulmonary insufficiency in patient's ehr.No echocardiogram was performed during admission, and there is not a previous echo available in the patient's ehr.Patient had lung cancer and had recently had chemotherapy.There was no report of malfunction for the conmed device.This report is being raised due to advisement of patient death.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device has been returned to conmed; however, the complaint investigation has not been completed.A supplemental and final report will be filed following the completion of the complaint investigation.We will continue to monitor per regulatory requirements to help ensure patient safety.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation of the returned device found that the gas connector was loose and missing the o-ring which may cause gas supply errors during use, however, the house gas connector is not used on this device.This device is used with bottle gas only and the house gas would not be used (no house gas in these surgery rooms).No evidence of mechanical failure, or calibration errors during preventative maintenance and subsequent evaluation.Air embolism is typically associated with gas entering the vascular system during surgical procedures.There is insufficient evidence to believe there was a malfunction of the airseal device that would / could have potentially caused the venous gas embolism.Possible causes include incorrect trocar placement, insufflation in non-sealed cavities, or deviation from operating guidelines.Preventative maintenance has been completed on this repair order.Repair/evaluation completed.Final test completed and met all specifications.A device history record (dhr) review was requested from the manufacturer, and no indication of abnormalities was communicated.The service history was reviewed and found similar repairs to this complaint.A two-year review of complaint history revealed there has been a total of 45 complaints, regarding 45 devices, for this device family and failure mode.(b)(4).To reduce the risk, use a low flow rate for the first insufflation and ensure that the insufflation instrument is correctly positioned.We will continue to monitor per regulatory requirements to help ensure patient safety.
|
| |
|
Event or Problem Description
|
|
The sales representative reported on behalf of the customer that the device, as-ifs1, airseal ifs, 120v, was being used during a right thoracic case on (b)(6) 2025 when it was reported, ¿embolism caused a stroke¿resulted in a patient death.¿.Further assessment was requested from the reporter, and it was found that the surgery was 2 hours and 26 minutes.Bottle gas was used, and the device pressure was set at 5.No history of sickle cell anemia, and there is no mention of pulmonary insufficiency in patient's ehr.No echocardiogram was performed during admission, and there is not a previous echo available in the patient's ehr.Patient had lung cancer and had recently had chemotherapy.There was no report of malfunction for the conmed device.This report is being raised due to advisement of patient death.
|
| |
|