• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CRITICAL CARE AB FLOW-I C20 ANESTHESIA SYSTEM; GAS-MACHINE, ANESTHESIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CRITICAL CARE AB FLOW-I C20 ANESTHESIA SYSTEM; GAS-MACHINE, ANESTHESIA Back to Search Results
Model Number 6888520
Medical Device Problem Code Failure of Device to Self-Test (2937)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/02/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Our investigation of the reported failure in this complaint has been completed.During the on-site assessment by the distributor¿s field service engineer (fse), the issue could be confirmed.The faulty reflector pressure transducer pcb was replaced which resolved the issue and the anesthesia system has since been returned to clinical use.However, as the device log was not obtained, a thorough evaluation could not be performed.Additionally, the replaced pressure transducer pcb was retained by the customer and was not returned, thereby preventing further root cause analysis.The pressure transducer pcb is part of the pressure measuring in the system.A faulty pressure transducer pcb may lead to inaccuracy in pressure measurement.This will be detected during system checkout and high priority alarms will be generated if the failure occurs during treatment.Our conclusion, based on the available information, is that the issue was resolved by replacing the faulty pressure transducer pcb.
 
Event or Problem Description
It was reported that the anesthesia system failed the pressure transducer test during system checkout.There was no patient involvement.Manufacturer's reference #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLOW-I C20 ANESTHESIA SYSTEM
Common Device Name
GAS-MACHINE, ANESTHESIA
Manufacturer (Section D)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer (Section G)
MAQUET CRITICAL CARE AB
roentgenvagen 2
solna
Manufacturer Contact
caroline kabbabe
roentgenvagen 2
solna 
MDR Report Key23876057
Report Number8010042-2025-0001899
Device Sequence Number11945104
Product Code BSZ
UDI-Device Identifier07325710010617
UDI-Public(01)07325710010617(11)231005
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K191027
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6888520
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/02/2025
Initial Report FDA Received Date12/23/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/05/2023
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-