• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WATCHMAN FLX? PRO; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WATCHMAN FLX? PRO; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL Back to Search Results
Model Number M635WU60400
Medical Device Problem Code Failure to Seal (4070)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/26/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Reported via literature article.It was reported that a device leak occurred.A left atrial appendage (laac) closure procedure was performed in 234 patients using previous-generation watchman flx (wmf) between may-21 and 14-feb-24, and in 134 patients using wmf-pro between 21-feb-24, and mar-2025.The background, procedural and short-term outcomes of consecutive 46 patients who underwent laac using large wmf-pro devices (35mm) and compared 35mm (33) with 40mm (13) was evaluated.This study was approved by the kurashiki central hospital irb and adhered to the declaration of helsinki.Device position, anchoring, seal, size, and release were confirmed by transesophageal echocardiography (tee) and fluoroscopy.Successful laac was defined as correct device placement with residual flow less than or equal to 5mm.Successful laac was achieved in 97 percent of wmf cases and 100 percent of wmf-pro cases.Follow-up tee or cardiac contrast (cct) was performed three (3) months post procedure.The imaging follow-up at 3 months were performed.There was no difference in the peri device leak (pdl) or patency at three (3) months were observed between the two groups, and no device related thrombus (drt) at 3 months was diagnosed in either group.However, pdl or gap was more frequently diagnosed in the wmf-pro 40mm group and significant pdl or gap greater than three to five millimeters was significantly more frequently diagnosed in this group.Significant pdls at 3 months persisted on the posterior-side.Among patients with significant pdl at 3 months, cases 1 and 3 had prominent pectinate muscles.No ischemic stroke events occurred in either group at 3 months.Previous study reported change in significant pdl greater than three to five millimeters between 45 days and one (1) year increased in one patient, unchanged in 42, and improved in 81 patients.In this report, two out of three pdls in the 40mm group were apparently associated with large pectinate muscles rather than with large ostial size itself.Therefore, such pdls are not logically linked to the larger device diameter.The transition from wmf to wmf-pro showed a similar proportion of large device usage.The 40mm device allowed closure of larger laas and demonstrated comparable procedural safety.However, use of 40mm device is technically more challenging and may be associated with residual pdl.
 
Additional Manufacturer Narrative
B3: estimated as 26/10/25 as the published date for the article is noted as october 26, 2025.D4 unique identifier (udi) #: detailed product information was not provided to bsc.Based on the nature of the information provided to bsc, it is not possible to perform a good faith effort to obtain the information.Because the product is unknown at this time, we are unable to provide the complete unique identifier (udi) # and other product specific information.If additional details become available, a supplemental report will be submitted.Literature citation: r.Chatani., h.Harata., m.Fujita., nishiura, n., mushiake, k., ono, s., tasaka, h., maruo, t., kadota, k., kubo, s.(2025).Potential of a novel large left atrial appendage closure device: short-term outcomes of the watchman flx pro 40-mm versus 35-mm.Catheterization and cardiovascular interventions 106 (2025): 3959-3962.Https://doi.Org/10.1002/ccd.70282.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WATCHMAN FLX? PRO
Common Device Name
SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key23877063
Report Number2124215-2025-92605
Device Sequence Number20026518
Product Code NGV
UDI-Device Identifier00191506004637
UDI-Public(01)00191506004637
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P130013
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Literature,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM635WU60400
Device Catalogue NumberM635WU60400
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/07/2025
Initial Report FDA Received Date12/23/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-