| Model Number |
261048 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Bone Fracture(s) (1870)
|
| Date of Event |
12/04/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
When inserting the 8 mm diameter manual reamer into the left humerus, the reamer broke at its thickest point, two-thirds of the way in, despite normal surgical use.An attempt to extract the piece of reamer resulted in a fracture of the humerus in the patient, leading to a change in the surgical indication, which was discussed with the head of the department.Consequences for patients: increased length of hospital stay / worsening health condition / rehospitalisation-reoperation-transfer / predictable complaints change in surgical indication, switching from total shoulder replacement to humeral nailing in a young patient.
|
| |
|
Event or Problem Description
|
|
When inserting the 8 mm diameter manual reamer into the left humerus, the reamer broke at its thickest point, two-thirds of the way in, despite normal surgical use.An attempt to extract the piece of reamer resulted in a fracture of the humerus in the patient, leading to a change in the surgical indication, which was discussed with the head of the department.Consequences for patients: increased length of hospital stay / worsening health condition / rehospitalisation-reoperation-transfer / predictable complaints change in surgical indication, switching from total shoulder replacement to humeral nailing in a young patient.
|
| |
|