• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FH INDUSTRIE ARROW; ARROW REAMER D8 L120

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FH INDUSTRIE ARROW; ARROW REAMER D8 L120 Back to Search Results
Model Number 261048
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Bone Fracture(s) (1870)
Date of Event 12/04/2025
Type of Reportable Event Serious Injury
Event or Problem Description
When inserting the 8 mm diameter manual reamer into the left humerus, the reamer broke at its thickest point, two-thirds of the way in, despite normal surgical use.An attempt to extract the piece of reamer resulted in a fracture of the humerus in the patient, leading to a change in the surgical indication, which was discussed with the head of the department.Consequences for patients: increased length of hospital stay / worsening health condition / rehospitalisation-reoperation-transfer / predictable complaints change in surgical indication, switching from total shoulder replacement to humeral nailing in a young patient.
 
Event or Problem Description
When inserting the 8 mm diameter manual reamer into the left humerus, the reamer broke at its thickest point, two-thirds of the way in, despite normal surgical use.An attempt to extract the piece of reamer resulted in a fracture of the humerus in the patient, leading to a change in the surgical indication, which was discussed with the head of the department.Consequences for patients: increased length of hospital stay / worsening health condition / rehospitalisation-reoperation-transfer / predictable complaints change in surgical indication, switching from total shoulder replacement to humeral nailing in a young patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW
Common Device Name
ARROW REAMER D8 L120
Manufacturer (Section D)
FH INDUSTRIE
6 rue nobel
quimper, 29000
FR  29000
Manufacturer (Section G)
FH INDUSTRIE
6 rue nobel
quimper, 29000
FR   29000
Manufacturer Contact
hind chouibani
6 rue nobel
quimper, 29000
FR   29000
MDR Report Key23877864
Report Number3003898228-2025-00005
Device Sequence Number18543013
Product Code KWS
UDI-Device Identifier03661489610486
UDI-Public03661489610486
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K093599
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other Health Care Professional
Remedial Action Modification/Adjustment
Type of Report Initial,Followup
Report Date (Section B) 02/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number261048
Device Lot NumberB03898
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/05/2025
Supplement Date Received by Manufacturer12/05/2025
Initial Report FDA Received Date12/23/2025
Supplement Report FDA Received Date02/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other;
Patient SexUnknown
-
-