VANTIVE US HEALTHCARE LLC REVACLEAR; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 114745
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/04/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
H11: the actual device was not available; however, photographs of the sample were provided for evaluation.Visual inspection of the photos showed that the filter dialysate port was broken where it connects to the dialyzer housing, which allowed the reported leakage.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the damage was not determined as no further or actual sample testing could be conducted.Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
E1: initial reporter city: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that an external fluid leak was observed from a "broken part" of a revaclear 300 during priming.There was no patient involvement.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVACLEAR
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
VANTIVE US HEALTHCARE LLC
round lake IL
Manufacturer (Section G)
VANTIVE HEALTHCARE - HECHINGEN BADEN DIALYZER
holger-crafoord-strasse 26
hechingen, baden-wurttemberg 72379
GM   72379
Manufacturer Contact
510 lake cook rd
deerfield, il, IL 60015
8153072966
MDR Report Key23878916
Report Number9611369-2025-00210
Device Sequence Number9622947
Product Code KDI
UDI-Device Identifier07332414114916
UDI-Public07332414114916
Combination Product (Y/N)N
Initial Reporter CountryFI
PMA/510(K) Number
K130039
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 01/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number114745
Device Lot Number5-9234-H01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/04/2025
Supplement Date Received by Manufacturer01/26/2026
Initial Report FDA Received Date12/23/2025
Supplement Report FDA Received Date01/29/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
NA.
Patient SexUnknown
-
-