| Catalog Number |
114745 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/04/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: the actual device was not available; however, photographs of the sample were provided for evaluation.Visual inspection of the photos showed that the filter dialysate port was broken where it connects to the dialyzer housing, which allowed the reported leakage.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the damage was not determined as no further or actual sample testing could be conducted.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
E1: initial reporter city: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that an external fluid leak was observed from a "broken part" of a revaclear 300 during priming.There was no patient involvement.No additional information is available.
|
| |
|