Catalog Number CLXSP-I
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/18/2025
Type of Reportable Event Malfunction
Event or Problem Description
The customer contacted therakos to report they experienced a pressure dome leak with their cellex photopheresis kit("kit") during an extracorporeal photopheresis (ecp) treatment.The customer reported they observed a leak at the pressure dome after 200mls of whole blood was processed.The treatment was aborted and no blood was returned to the patient.The patient was reported to be stable.The kit return was requested; however, the customer discarded the kit.The customer will not return product for investigation.
 
Additional Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction pressure dome leak.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot p123 was conducted.There were no non-conformances related to the complaint.This lot met all release requirements.A review of kit lot p123 shows no trends.Trends were reviewed for complaint category, pressure dome leak.No trends were detected for this complaint category.The assessment is based on the information available at the time of investigation.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Complaints are monitored through tracking and trending.If a trend is detected, further investigation will be conducted.No further action is required at this time.This investigation is now complete.(b)(4).(b)(6) 2025.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Common Device Name
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
THERAKOS DEVELOPMENT LIMITED
blanchardstown, dublin D15 A P2D
EI  D15 AP2D
Manufacturer (Section G)
HARMAC MEDICAL PRODUCTS
i.d.a. business park
castlerea roscommon,
EI  
Manufacturer Contact
mark wendelken
440 route 22 east
suite 302
bridgewater, NJ 08807
MDR Report Key23879337
Report Number3036773437-2025-00029
Device Sequence Number9605160
Product Code LNR
UDI-Device Identifier00705030000005
UDI-Public(01)00705030000005(10)P123(17)270301
Combination Product (Y/N)Y
Initial Reporter CountryGM
PMA/510(K) Number
P860003
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 12/23/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberCLXSP-I
Device Lot NumberP123
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/18/2025
Initial Report FDA Received Date12/23/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/29/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-