Model Number 6944-58
Medical Device Problem Codes Fracture (1260); High impedance (1291); Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/14/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the right ventricular (rv) lead triggered warnings for high rv defibrillation, high superior vena cava (svc), and high rv pacing impedances.After further review, a rise over time eventually resulting in a sudden increase of the rv defibrillation and svc coil impedances was noted.In addition, a ride in the rv pacing impedances over time was noted.A fracture was suspected.The rv lead was capped and replaced.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Continuation of d10: 5076-45 lead implanted: (b)(6) 2010 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated the impedance of the right ventricular defibrillation coil was beyond the expected upper range.Analysis of the device memory indicated the impedance trend of the right ventricular defibrillation coil was rising.Analysis of the device memory indicated the impedance of the right ventricular defibrillation coil had a spike.Analysis of the device memory indicated the impedance of the superior vena cava defibrillation coil was beyond the expected upper range.Analysis of the device memory indicated the impedance trend of the superior vena cava defibrillation coil was rising.Analysis of the device memory indicated the impedance of the superior vena cava defibrillation coil had a spike.Analysis of the device memory indicated the impedance of the right ventricular pacing lead was beyond the expected upper range.Analysis of the device memory indicated the impedance trend of the right ventricular pacing lead was rising.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPRINT QUATTRO LEAD
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (NON-CRT)
Manufacturer (Section D)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MEDTRONIC PUERTO RICO VILLALBA
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
gerard torenvliet
8200 coral sea st ne
mounds view, MN 55112-4391
7635269277
MDR Report Key23879529
Report Number2649622-2025-35862
Device Sequence Number4771781
Product Code LWS
UDI-Device Identifier00613994249951
UDI-Public(01)00613994249951(17)110826
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P920015
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/12/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date08/26/2011
Device Model Number6944-58
Device Catalogue Number6944-58
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/09/2025
Supplement Date Received by Manufacturer12/30/2025
Initial Report FDA Received Date12/23/2025
Supplement Report FDA Received Date01/12/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/03/2009
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DDMB1D1 ICD
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age69 YR
Patient SexFemale
Patient Weight90 KG
-
-