| Model Number |
PIN DRIVER ATTACHMENT, 0-3.2MM |
| Medical Device Problem Codes |
Material Fragmentation (1261); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/02/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Event or Problem Description
|
|
It was reported that during a foot surgery the ar-400 handpiece got very hot, and the k-wire didn't appear to rotate normally and created a sort-of metal "dust" and even small metal "flakes".It clearly jammed or obstructed the k-wire from normal rotation.After normal troubleshooting, the surgeon switch to another ar-400pd, and everything ran smoothly.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.The manufacturer evaluation revealed a broken bearing at the drive shaft as well as a sticking clamping system.
|
| |
|