MAUDE Adverse Event Report: MENTOR TEXAS ARTOURA BREAST TISSUE EXPANDER; EXPANDER, SKIN, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS ARTOURA BREAST TISSUE EXPANDER; EXPANDER, SKIN, INFLATABLE Back to Search Results
Catalog Number TEXP155RH
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code Deformity/ Disfigurement (2360)
Date of Event 11/02/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
Device evaluation completed on april 06, 2026: the product was returned to mentor for evaluation.Mentor then conducted a visual inspection, leak testing, and microscopic examination of the returned device.Visual analysis of the returned sample revealed that no damage or anomalies were observed on the tissue expander.Leak testing was performed, in accordance with mentor procedures, and a tear was observed at the junction between the shell and base at the nine o'clock position.No additional leak sites.Microscopic examination was performed and the cause of the tear could not be identified.Mentor also performed a manufacturing record evaluation related to the reported complaint for the finished device lot number.No manufacturing non-conformances were identified as part of this evaluation.Based on the information currently available, a notification was sent to mentor's manufacturing team for further investigation.In conclusion, according to the evaluation performed and data records reviewed on the complaint, the deflation reported is not able to be confirmed as a manufacturing defect.Although no conclusion could be reached on the cause of the reported event, the instructions for use contain the following caution: the silicone shell of the tissue expander may easily be cut by a scalpel or ruptured by excessive stress, manipulation with blunt instruments, or penetration by a needle.Subsequent deflation and/or rupture could result.All devices should be carefully inspected for structural integrity prior to and during implantation.In addition, the patient should be advised that vigorous body movement (e.G., physical exercise) or excessive manipulation or trauma in the expander region may cause stress to the device and result in subsequent deflation.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria: no corrective and preventive action (capa) is required now.A review of the manufacturing records was conducted to determine whether lot number 2055391 was associated with any nonconformances and to identify any in process device rejections for defects similar to the tear reported in the complaint.The record review confirmed that no devices exhibited- the same failure mode as the reported complaint.Additionally, no nonconformances were recorded for this lot.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
A female patient who underwent a breast reconstruction with mentor artoura breast tissue expander, 850cc saline prosthesis experienced left sided deflation post operation.As a result, the patient underwent unilateral replacement with lot: 2055391 on (b)(6) 2025.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTOURA BREAST TISSUE EXPANDER
Common Device Name
EXPANDER, SKIN, INFLATABLE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
9497882078
MDR Report Key23881351
Report Number1645337-2025-14197
Device Sequence Number20010314
Product Code LCJ
UDI-Device Identifier00081317009511
UDI-Public(01)00081317009511(11)250315(17)290314(10)2055391
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K150777
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberTEXP155RH
Device Lot Number2055391
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 12/02/2025
Supplement Date Received by Manufacturer01/12/2026
04/06/2026
Initial Report FDA Received Date12/23/2025
Supplement Report FDA Received Date02/02/2026
04/07/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured03/15/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age47 YR