• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CENTURION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065752182
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/26/2025
Type of Reportable Event Malfunction
Event or Problem Description
A physician reported that while using ultrasound mode, abnormal noises was heard from the part of the cassette pack during the cataract with intraocular lens implant surgery.The surgery was completed on the same day but replacement details were unknown.There was no information about patient impact.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Additional information provided in d.9., h.3., h.6.And h.11.One turbid active fluid management system (fms) in the tray was visually inspected.The drain bag was detached from the drain bag tape on the cover due to saturation.The drain bag tape was adhered on the cassettes properly.Both irrigation and aspiration luers were intact.Flare shape was observed in the blue socket fitting at the aspiration tubing insertion end.The product was tested using the centurion console with the current software.The product could be recognized by the console.The product primed and tuned with the sentry handpiece and the 0.9mm aspiration bypass system (abs) tip and infusion sleeve from the lab stock, successfully and the service data could be retrieved.No system message code displayed on console screen.The aspiration flow rate and irrigation free flow met specification.Neither leakage, nor occlusion was detected from the tubing insertion to the fittings and the four aspiration ports on the pump region of the cassette base.The products functioned within specification.The investigation conducted a non-conformance review of the reported lot number.No deviations that could have contributed to this event were identified, all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event.The product functioned per specifications.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.No further investigative or manufacturing actions are warranted at this time due to the product meeting specifications.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all company products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SURGICAL PROCEDURE PAK
Common Device Name
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23881859
Report Number1644019-2025-05686
Device Sequence Number4706930
Product Code LRO
UDI-Device Identifier00380657521821
UDI-Public(01)00380657521821(17)270331(11)250403(10)17C5HP
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K880961
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/01/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8065752182
Device Lot Number17C5HP
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/27/2025
Supplement Date Received by Manufacturer03/11/2026
Initial Report FDA Received Date12/23/2025
Supplement Report FDA Received Date04/01/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/03/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-