| Model Number |
M0054243R0 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Optical Problem (3001)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Block b3: approximated based on the date the manufacturer became aware of the event.Block h6: imdrf code a06 captures the reportable event of loss of visualization.
|
| |
|
Event or Problem Description
|
|
Note: this report pertains to one of two devices used in the same procedure.It was reported to boston scientific corporation that an exalt model d scope was used with an exalt controller during an ercp (endoscopic retrograde cholangiopancreatography) procedure performed on an unknown date.During the procedure, the scope malfunctioned and the image went blank.The procedure was completed with a reusable scope.There was no patient complications reported as a result of this event.
|
| |
|
Event or Problem Description
|
|
Note: this report pertains to one of two devices used in the same procedure.It was reported to boston scientific corporation that an exalt model d scope was used with an exalt controller during an ercp (endoscopic retrograde cholangiopancreatography) procedure performed on an unknown date.During the procedure, the scope malfunctioned and the image went blank.The procedure was completed with a reusable scope.There was no patient complications reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Block b3: date of event
approximated based on the date the manufacturer became aware of the event.Block h6:
imdrf code a06 captures the reportable event of loss of visualization.Block h11: investigation summary.The exalt model d controller was returned for analysis.A product analysis identified moderate wear and tear to exterior however the controller passed the initial testing.Therefore, the reported allegation of loss of visualization in this complaint was not confirmed.A similar complaint review was completed.There were no emerging trends identified that warrant further action.The device history record review confirmed that the device met all material, assembly, and performance specifications prior to distribution.It was confirmed this device met manufacturing specification prior to distribution and there were no manufacturing deviations which could have contributed to the reported event.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of no problem detected.
|
| |
|