Model Number 1100403
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/16/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging an issue related to a simplygo device involving a thermal product malfunction.There were no reports of patient harm, injury, or medical intervention.The device was returned to a third-party service center, where it was found that the unit did not start and the pca was burned.As per customer request, the unit was required to be evaluated at riql and shipped unrepaired.Due to the alleged malfunction, the device was forwarded to the manufacturer¿s product investigation lab (pil) for additional testing and investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMPLYGO
Common Device Name
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
Respironics, Inc.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
+1(724)733-0200
MDR Report Key23882680
Report Number2518422-2025-112997
Device Sequence Number9464333
Product Code CAW
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K111885
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation 505
Type of Report Initial,Followup
Report Date (Section B) 09/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1100403
Device Catalogue Number1100403
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/31/2025
Supplement Date Received by Manufacturer07/23/2025
Initial Report FDA Received Date12/23/2025
Supplement Report FDA Received Date09/15/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/01/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device N
Patient Sequence Number1
Patient SexUnknown
-
-