MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION FLEXIVA PULSE ID; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION FLEXIVA PULSE ID; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number M006L8406940
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/03/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during procedure, the tip of the laser fiber broke at the exposed glass tip.The same event occurred with two fibers.The procedure was completed using another of the same device.There were no patient complications.This event is for the second fiber.
 
Additional Manufacturer Narrative
The device is not available for analysis; therefore, no physical analysis of the product could be performed.The product record review results detailed in the prr table did not determine probable cause.A review of the device history record (dhr) confirmed that the device met specification prior to shipping.There was no manufacturing deviations noted that could have contributed to the event reported.A labeling review was performed on the device instructions for use (ifu).The ifu cited that, care should be exercised with the glass tip to avoid severe impact or side stresses that may fracture the tip.Based on the information available, the conclusion code "unable to exclude device problem" has been assigned as the most probable cause.
 
Event or Problem Description
It was reported that during procedure, the tip of the laser fiber broke at the exposed glass tip.The same event occurred with two fibers.The procedure was completed using another of the same device.There were no patient complications.This event is for the second fiber.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIVA PULSE ID
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
pmt 741 persiaran cassia selat
bandarcassia, pulau pinan 14110
MY   14110
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key23882797
Report Number2124215-2025-92798
Device Sequence Number17326488
Product Code GEX
UDI-Device Identifier08714729978701
UDI-Public(01)08714729978701(17)280604(10)36682072
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
K210925
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/17/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberM006L8406940
Device Catalogue NumberM006L8406940
Device Lot Number0036682072
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/03/2025
Supplement Date Received by Manufacturer01/27/2026
Initial Report FDA Received Date12/23/2025
Supplement Report FDA Received Date02/17/2026
Date Device Manufactured06/05/2025
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-