H7) this regulatory report is being submitted due to retrospective review through fa1523, 806 report # 3005075696-12/16/2025-001-c.Continuation of d10: section d information references the main component of the system.Other relevant device(s) are: product id: kit1378-06, software version: 5.1.2 h3, h6) a software analysis was performed for this event.In summary, the complaint was confirmed and a software issue was identified.Codes b01, c10, and d18 are applicable.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Medtronic received information regarding a guidance system being used during a spinal procedure.It was reported that the system became unresponsive during a scan and plan procedure.The issue occurred when attempting to upload the medtronic imaging scan via a universal serial bus (usb), leading to a delay of about five minutes.The team decided not to attempt reloading the scan to avoid further risk and instead performed a second scan, which loaded without issues.There was no impact on patient outcome.
|