Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
It was reported that during use, a low tidal volume alarm was triggered and the preset value could not be reached.The vital signs did not change much, but the blood oxygen was a little low for about 20 minutes.An urgent inspection of the ventilator and the entire circuit revealed an internal crack at the inner layer of the double interface of the breathing circuit.The device was immediately replaced.
|