Additional information provided in d.9., h.3., h.6.And h.11.One opened probe with tip protector in a bubble bag was received.The sample was visually inspected and found to be nonconforming with the probe needle bent at stiffener sleeve and orange/brown foreign material on the port face and welded cap.The sample was then functionally tested for actuation and aspiration.The actuation and aspiration functionalities of the returned probe were both nonconforming.No wear assessment could be performed due to the inner cutter broke while trying to extract it from the bent needle.A review of the device history record traceable to the lot number obtained from the device¿s radio frequency identification tag, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The investigation conducted a non-conformance review of the reported lot number.No deviations were identified, and all corresponding production release specifications defined in the device master record were met.Based on the evaluation of the information and materials received, the investigation was unable to identify the root cause or origin of the reported event.Additionally, no manufacturing-related deficiencies were found that potentially could have contributed to the complaint.The complaint evaluation confirmed the reported aspiration failure.The most likely cause for the aspiration and actuation failure is from the bent needle.A bent needle can impede the movement of the cutter shaft and aspiration flow through the probe.This interference is present as observed from visual condition of the inner cutter.The exact root cause of the bent needle cannot be determined from this evaluation.The most likely root cause is handling at any point after the probe was shipped from the manufacturing site including use during surgery.As the root cause and its origin are inconclusive, further investigative or manufacturing actions are not warranted at this time as the reported aspiration failure was confirmed, and an exact root cause for the bent needle could not be determined from this evaluation.All probes are 100% visually inspected and tested for actuation, aspiration, and cut during manufacturing.Any non-conformances found are removed from the lot and scrapped.Receipt of additional relevant information or supporting materials will prompt a re-evaluation of the complaint investigation.Complaint data for all products is reviewed monthly to monitor for adverse trends.During the last review, no adverse trends were observed for the reported product and event combination.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.No further action warranted at this time.The manufacturer internal reference number is: (b)(4).
|