ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM; INTRAOCULAR LENS Back to Search Results
Model Number AU00T0
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 12/09/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for evaluation.A photo was provided.The photo was of the implanted lens in the eye.There appeared to be a scratch mark off center of the optic, possibly on the anterior surface.The clarity of the photo did not allow for a determination.The observed scratch may indicate a plunger override occurred.A device history record review and a non-conformance review of the reported lot number was conducted.The deviation review did not reveal any potential contributing factors to the reported complaint and all corresponding production release specifications defined in the device master record were met.A qualified viscoelastic was indicated.The root cause for the reported event could not be determined.The company device was indicated not available for return by the reporting facility.Based on review of the provided photo, there appeared to be a scratch mark off center of the optic, possibly on the anterior surface.The observed scratch may indicate a plunger override occurred.The ifu instructs: fully insert the ovd cannula, perpendicular to the device, through the viscoelastic port, located in the lens stop portion of the device.Fill the device until ovd can be observed flowing to the ¿fill-to¿ line on the nozzle tip.This will require approximately 0.2 ml of ovd.Only use company ovd qualified for use with the company¿ pre-loaded delivery system that has been allowed to come to the operating room temperature.If inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly or become ¿stuck¿ in the device.The ifu instructs: take at least 7 seconds to gently fold the iol by advancing the plunger forward in one smooth, continuous motion until the front edge of the optic is even with the ¿fill-to¿ line on the nozzle.Do not allow any portion of the lens to exit the nozzle tip.At this position the blue spring will contact the clear main body.It is important to not advance the plunger abruptly to fold the iol as improper folding and lens damage may occur.The ifu instructs: after the lens has been advanced to the nozzle line, the lens should be visually inspected to determine the position of the haptics.The plunger should be in contact with the trailing optic edge.After confirming the lens is properly positioned and the haptics are folded properly, proceed with lens implantation.Proceeding with implantation of a misfolded haptic or a lens that appears to be ¿out of position¿ can result in a broken haptic or other negative outcome, since the haptic may be trapped and stretched, and/or pinched and sheared by the moving plunger.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A healthcare professional reported that during an intraocular lens (iol) implantation procedure, scratches or blemishes¿such as indentations, scuff marks, or score marks¿were observed on the optic after the lens was implanted.The procedure was completed on the same day using a backup lens.Additional information was requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ SINGLEPIECE IOL WITH ULTRASERT DELIVERY SYSTEM
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
246 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23887773
Report Number1119421-2025-03230
Device Sequence Number4669464
Product Code HQL
UDI-Device Identifier00380652394901
UDI-Public(01)00380652394901(17)280402(11)250404(10)16022101
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 12/24/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAU00T0
Device Lot Number16022101
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/11/2025
Initial Report FDA Received Date12/24/2025
Date Device Manufactured04/04/2025
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PROVISC OVD.
Patient SexUnknown
-
-