| Catalog Number |
0119L22 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/15/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that this was an incident of spontaneous catheter dislodgement.A 3-way 22fr foley catheter was placed for a patient with a thrombus.It had spontaneously dislodged after three days.It was unclear whether it came out due to sterile water leakage or if it burst after being removed while the balloon was inflated.The actual device has been discarded and could not be verified.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that this was an incident of spontaneous catheter dislodgement.A 3-way 22fr foley catheter was placed for a patient with a thrombus.It had spontaneously dislodged after three days.It was unclear whether it came out due to sterile water leakage or if it burst after being removed while the balloon was inflated.The actual device has been discarded and could not be verified.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because no sample was returned.It is unknown whether the device had met relevant specifications.It was unknown whether the product had caused the reported failure.The potential root cause for this failure could be due to thin sac causing by short latex dwell time, latex dip speeds out too slow and also might be contributed by user related (example: mishandling product or exposed to petroleum or sharp object).The labeling and instructions for use were found to be adequate.The dhr review could not be performed without a lot number.Therefore, no additional action is required at this time.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Search Alerts/Recalls
|