| Model Number |
SN-3697 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: sn (b)(6), monosof* 6-0 blk 45cm p10 x12, (lot#d4d2909fy); sn (b)(6), monosof* 6-0 blk 45cm p10 x12 (lot#d4d2909fy); sn (b)(6) monosof* 6-0 blk 45cm p10 x12 (lot#d4d2909fy); sn (b)(6), monosof* 6-0 blk 45cm p10 x12 (lot#d4d3259fy); sn (b)(6), monosof* 6-0 blk 45cm p10 x12 (lot#d4d3259fy);sn (b)(6), monosof* 6-0 blk 45cm p10 x12 (lot#d4d3259fy); sn (b)(6),, monosof* 6-0 blk 45cm p10 x12 (lot#d4d3259fy); sn (b)(6), monosof* 6-0 blk 45cm p10 x12 (lot#d4d3259fy); sn (b)(6), monosof* 6-0 blk 45cm p10 x12 (lot#d4d3259fy); sn (b)(6), monosof* 6-0 blk 45cm p10 x12 (lot#d4d3259fy); sn (b)(6), monosof* 6-0 blk 45cm p10 x12 (lot#d4d3259fy); sn (b)(6), monosof* 6-0 blk 45cm p10 x12 (lot#d4d3259fy); medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a procedure, the thread of 15 sutures broke in the middle, while other 9 sutures from different batch had the same issue.A new suture from a different batch was used to complete the case.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.One opened device with the appropriate packaging and eleven sealed representative samples were available for evaluation.The evaluation found no potentially contributing factors, and the sample met all related specifications.It was reported that sutures broke in the middle.The reported issue could not be confirmed.The most likely cause could not be identified because no related problem was detected with the device.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|