| Model Number |
SN-3697 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
12/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
D10 concomitant products: sn: (b)(6) monosof* 6-0 blk 45cm p10 x12 - (lot#: d4d2909fy); sn: (b)(6) monosof* 6-0 blk 45cm p10 x12 - (lot#: d4d2909fy); sn: (b)(6) monosof* 6-0 blk 45cm p10 x12 - (lot#: d4d2909fy); sn: (b)(6) monosof* 6-0 blk 45cm p10 x12 - (lot#: d4d2909fy); sn: (b)(6) monosof* 6-0 blk 45cm p10 x12 - (lot#: d4d3259fy); sn: (b)(6) monosof* 6-0 blk 45cm p10 x12 - (lot#: d4d3259fy); sn: (b)(6) monosof* 6-0 blk 45cm p10 x12 - (lot#: d4d3259fy); sn: (b)(6) monosof* 6-0 blk 45cm p10 x12 - (lot#: d4d3259fy); sn: (b)(6) monosof* 6-0 blk 45cm p10 x12 - (lot#: d4d3259fy); sn: (b)(6) monosof* 6-0 blk 45cm p10 x12 - (lot#: d4d3259fy); sn: (b)(6) monosof* 6-0 blk 45cm p10 x12 - (lot#: d4d3259fy); sn: (b)(6) monosof* 6-0 blk 45cm p10 x12 - (lot#: d4d3259fy).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a procedure, the thread of 15 sutures broke in the middle, while other 9 sutures from different batch had the same issue.A new suture from a different batch was used to complete the case.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.One opened device with the appropriate packaging and eleven representative sealed samples were available for evaluation.Visual inspection of the opened sample revealed two sutures and two needles.One opened breather pouch and an empty suture retainer were also returned.The needles remained attached to their sutures, which were broken.Suture "a" and suture "b" were broken from the needle attachment point.The measurements were taken with an aluminum rule.Inspection of the pouch showed no damage or breaches.The needles were properly parked in the retainer, and the sutures were properly wound with no damage to the retainer.A tactile and microscopic inspection of the sutures was performed with no abnormalities.Functional testing noted that the breathable pouch was opened without any tears of the tyvek packaging.The sutures were removed from the retainers without any difficulty.A simple knot was performed on the suture and the knot was pulled tight.The suture ends were loaded onto an instron machine by clamping the ends with cord yarn grips.No suture breaking or fraying was noted while hand ling or loading the suture into the instron machine.The instron then pulled the suture until the suture broke.The breaking point load force was measured and were found to meet specifications.Based on the results of the simple knot test, the representative samples tested satisfactory.It was reported that the sutures broke in the middle.The reported issue was confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|