| Catalog Number |
00598801013 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Pain (1994); Joint Laxity (4526)
|
| Date of Event |
12/09/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient with a history of unknown right total knee arthroplasty infection had a second stage revision performed.Subsequently, the patient was revised approximately 6 years post implantation due to instability.No further information has been provided at this time.Attempts have been made and additional information on the reported event is unavailable at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).D10: st prc tib plt size 4, #item 00598003702, #lot j6537701.Prc agmt block dist sz f 5mm, #item 00599003610, #lot 63974685.Prc agmt block dist sz f 5mm, #item 00599003610, #lot 63832559.Stem implant, #item 00598801012, #lot 64358368.Lcck fem implant sz f-r, #item 00599401692, #lot 64179221.Ng lcck art sf, #item 00599403212, #lot 63447463.All poly pat comp, #item 00597206532, #lot 64358540.Prc tib block 5mm sz4, #item 00-5988-004-26, #lot 63245817.Prc tib block 5mm sz4, #item 00-5988-004-26, #lot 64146401.Therapy date: (b)(6) 2025.G2: foreign source country: australia.H6 suggested component code: mechanical (g04) - stem.An investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: b4,b5,d2,g1,g3,g6,h1,h2,h3,h6.The cmp was not confirmed.- visual examination of the provided pictures identified explanted device with bone cement and foreign debris on the surfaces.As the implants were not returned, further evaluations could not be provided.- review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.- the reported products were reviewed for compatibility with no issues noted.- for the articular surface, femoral augments, 13mm diameter stem, and tibial augments: review of complaint history found no additional related issues for these items and the reported part and lot combinations.- radiographs were provided and reviewed by a health care professional and it was determined further review would not enhance the investigation.- medical records were provided and reviewed by a health care professional.Review of the available records identified the following: initial / stage 2 revision op note demonstrate no intraop complications.- a definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No further information is available at the time of this report.
|
| |
|
Search Alerts/Recalls
|
|