| Catalog Number |
550300500 |
| Medical Device Problem Code |
Osseointegration Problem (3003)
|
| Health Effect - Clinical Code |
Inadequate Osseointegration (2646)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H11: additional narrative: a1: patient identifier: (b)(6).D4: udi: as the lot number for the device involved in the event provided 232002 was not able to be validated, the full udi is currently not available.
|
| |
|
Event or Problem Description
|
|
Subject id: (b)(6).Study no: (b)(6).Clinical adverse event received for shoulder revision device and procedure (relatedness) device related: no information provided.Procedure related: no information provided.Date of event: no information provided.Date of implant: (b)(6) 2025.Date of revision: no.Information provided.Device location: no information provided.Treatment/impact: were depuy synthes components removed: yes.Clinical adverse event received for revision ¿ aseptic loosening.Date of implant: (b)(6) 2025.Date of revision: (b)(6) 2025.Device location: right.Treatment/impact: were depuy synthes components removed: yes (cup, metaglene, head/glenosphere); enhance reverse shoulder arthroplasty was revised and converted to an enhance hemiarthroplasty, for treatment of aseptic loosening of the enhance baseplate and screws.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform as part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
| |
|
Search Alerts/Recalls
|