| Catalog Number |
4825202 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
10/29/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided.
|
| |
|
Event or Problem Description
|
|
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the red blood cell (rbc) product.Donor unit id: (b)(6) there was not a transfusion recipient or patient involved at the time of the unit processing; therefore, no patient information is reasonably known at the time of the event.The collection set is not available for return because it was discarded by the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being filed to provide additional information in h.6 and h.11.Investigation: the device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.Investigation is in process.A follow-up report will be provided.
|
| |
|
Event or Problem Description
|
|
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the red blood cell (rbc) product.Donor unit id: (b)(6) there was not a transfusion recipient or patient involved at the time of the unit processing; therefore, no patient information is reasonably known at the time of the event.The collection set is not available for return because it was discarded by the customer.
|
| |
|
Event or Problem Description
|
|
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the red blood cell (rbc) product.Donor unit id: (b)(6) there was not a transfusion recipient or patient involved at the time of the unit processing; therefore no patient information is reasonably known at the time of the event.The collection set is not available for return because it was discarded by the customer.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being filed to provide additional information in d.4, h.6 and h.11.Investigation: the device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.The run data file (rdf) was analyzed for this event.The analysis of the run data file did not identify a conclusive root cause for the higher-than-expected wbc content reported in the rbc product for this collection.No unusual process variable was identified from the signals in the run data file.Based on the available information, it cannot be ruled out that a problem with the rbc filter in the specific tubing set used for this collection may have contributed to the leukoreduction failure.Additionally, it also cannot be ruled out that the higher-than-expected wbc content in the rbc product may have been donor-related.Root cause: the analysis of the run data file did not identify a conclusive root cause for the higher-than-expected wbc content reported in the rbc product for this collection.No unusual process variable was identified from the signals in the run data file.Based on the available information, it cannot be ruled out that a problem with the rbc filter in the specific tubing set used for this collection may have contributed to the leukoreduction failure.Additionally, it also cannot be ruled out that the higher-than-expected wbc content in the rbc product may have been donor-related.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being filed to provide additional information in h.6 and h.11.Investigation: the device history records (dhr) were reviewed for this lot.There were no events noted in the dhr that would have contributed to the elevated wbc count experienced by the customer.The run data file (rdf) was analyzed for this event.The analysis of the run data file did not identify a conclusive root cause for the higher-than-expected wbc content reported in the rbc product for this collection.No unusual process variable was identified from the signals in the run data file.Based on the available information, it cannot be ruled out that a problem with the rbc filter in the specific tubing set used for this collection may have contributed to the leukoreduction failure.Additionally, it also cannot be ruled out that the higher-than-expected wbc content in the rbc product may have been donor-related.Investigation is in process.A follow-up report will be provided.
|
| |
|
Event or Problem Description
|
|
The customer would like the run data file investigated to determine a possible cause for the elevated white blood cell (wbc) content in the red blood cell (rbc) product.Donor unit id: (b)(6).There was not a transfusion recipient or patient involved at the time of the unit processing, therefore no patient information is reasonably known at the time of the event.The collection set is not available for return because it was discarded by the customer.
|
| |
|