| Model Number |
G150 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/04/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during a procedure to replace the right ventricular (rv) lead in this cardiac resynchronization therapy defibrillator (crt-d) system, the right atrial (ra) lead and crt-d were explanted and replaced due to high pacing impedance measurements.At this time, this crt-d has not been returned for analysis, and no additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device technical analysis: this product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent changes in impedance measurements with no conclusive evidence of a product performance issue or inadequate lead-to-device connection; please refer to the event description for more information regarding the specific circumstances of this event.Engineering analysis and testing of returned products has determined that repeated, small movements of the lead terminal ring along with variances in the connection contact force and surface roughness can impact the connection between the spring contact and the lead ring, resulting in intermittent changes in impedance measurement.Device history record review: a review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.Labeling review: review of labeling determined that the complaint situation was listed in the manual.There was no indication in the complaint that the product was not used in accordance with labeling.The manual was unlikely to be the cause of the reported complaint; translation, wording, or graphics does not require further review.Risk assessment: a risk review was completed and confirmed that the event of shock impedance high out of range - increasing gradually was defined in the risk documentation.This event type has been accounted for during product risk analysis to support acceptable risk benefit for the product.Investigation conclusion: based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of cause traced to device design.If pertinent information is provided in the future, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that during a procedure to replace the right ventricular (rv) lead in this cardiac resynchronization therapy defibrillator (crt-d) system, the right atrial (ra) lead and crt-d were explanted and replaced due to high pacing impedance measurements.At this time, this crt-d has not been returned for analysis, and no additional adverse patient effects were reported.
|
| |
|